Oral Abstracts Session
Virtual Recording
J. Jane J. Cao, MD, MPH
Director of Translational Research, System director of cardiac imaging
St. Francis Hospital, The Heart Center
Roslyn, New York, United States
J. Jane J. Cao, MD, MPH
Director of Translational Research, System director of cardiac imaging
St. Francis Hospital, The Heart Center
Roslyn, New York, United States
Jonathan Weber, MPH
Research Analyst
St. Francis Hospital, The Heart Center
Greenvale, New York, United States
Michael Passick, MBA, RT
Manager, Research Echo
St. Francis Hospital, The Heart Center, New York, United States
Nora Ngai, PhD
Core lab coordinator
St. Francis Hospital, The Heart Center, United States
Karli PIpitone, MPH
Research Analyst
St. Francis Hospital & Heart Center
Huntington, New York, United States
Kristine Estrin
Clinical Research Echosonographer
St. Francis Hospital, The Heart Center, United States
RV Measurement | Method | Mean 1 | Mean 2 | Mean difference | ICC | MDD |
End diastolic volume | Automatic | 148.6 ± 37.8 | 150.9 ± 40.6 | -2.3 ± 16 | 0.92 (0.79, 0.97) | 31.4 |
Auto + adjustment | 146.2 ± 39.5 | 150 ± 37.7 | -3.8 ± 13.1 | 0.92 (0.8, 0.97) | 29.3 | |
Manual | 149.4 ± 45.9 | 149.8 ± 40.8 | -0.3 ± 11 | 0.97 (0.92, 0.99) | 21.6 | |
End systolic volume | Automatic | 65.9 ± 25.2 | 71.9 ± 27.4 | -6 ± 7.8 | 0.94 (0.71, 0.98) | 15.2 |
Auto + adjustment | 60.3 ± 22.3 | 62.6 ± 23.2 | -2.3 ± 6.2 | 0.95 (0.86, 0.98) | 12.8 | |
Manual | 59.9 ± 23.5 | 62.1 ± 22.6 | -2.2 ± 8 | 0.94 (0.84, 0.98) | 15.8 | |
Stroke volume | Automatic | 82.7 ± 18.8 | 79 ± 17.6 | 3.7 ± 15.4 | 0.64 (0.25, 0.86) | 30.1 |
Auto + adjustment | 85.9 ± 19.6 | 87.4 ± 16.1 | -1.5 ± 9.8 | 0.82 (0.57, 0.93) | 23 | |
Manual | 88.4 ± 24.8 | 87.6 ± 20.3 | 0.8 ± 6.9 | 0.96 (0.88, 0.98) | 13.5 | |
Ejection fraction | Automatic | 56.4 ± 7.7 | 53.2 ± 6.8 | 3.3 ± 6.1 | 0.60 (0.18, 0.84) | 11.9 |
Auto + adjustment | 59.5 ± 5.1 | 59.2 ± 5 | 0.3 ± 3 | 0.79 (0.49, 0.92) | 6.7 | |
| Manual | 59.7 ± 5.3 | 59.2 ± 4.8 | 0.5 ± 1.8 | 0.94 (0.83, 0.98) | 3.5 |
ICC = inter-class correlation coefficient; MDD = minimum detectable differenceTable 2. Detectable changes in RV volumes in serial studies among cardiomyopathy and cardio-oncology cohorts
RV Measurement | Change between serial studies | Detectable changes by post-processing method | ||||
Mean | Min | Max | Automatic | Auto + adjustment | Manual | |
Cardiomyopathy (N=95) | ||||||
End diastolic volume | 1.1 ± 28 | -51 | 83 | 22 (23%) | 30 (32%) | 43 (45%) |
End systolic volume | 1.3 ± 14.6 | -27.2 | 55.3 | 21 (22%) | 31 (33%) | 19 (20%) |
Stroke volume | -0.3 ± 19.7 | -50.5 | 60.4 | 13 (14%) | 25 (26%) | 42 (44%) |
Ejection fraction | -0.5 ± 6.7 | -24 | 14 | 6 (6%) | 29 (31%) | 58 (61%) |
Cardio-oncology group (N=24) | ||||||
End diastolic volume | 7.6 ± 27.9 | -51 | 83 | 7 (29%) | 8 (33%) | 12 (50%) |
End systolic volume | 3.1 ± 11.9 | -27.2 | 55.3 | 2 (8%) | 7 (29%) | 2 (8%) |
Stroke volume | 4.6 ± 22.8 | -50.5 | 60.4 | 5 (21%) | 5 (21%) | 11 (46%) |
Ejection fraction | -0.5 ± 6.9 | -24 | 14 | 3 (13%) | 8 (33%) | 14 (58%) |